Pulse Rate selection: in the continuous mode 2 Hz – 200 Hz [2% accuracy].
Mode: Continuous, Burst or Modulated.
Burst mode: Bursts of 9 pulses [200 µS] at 150 Hz, repeating twice every second.
Modulation mode: 6-second cycle of concurrent width modulation and pulse repetition rate modulation.
Width starting at 200 µS and decreasing exponentially to 100 µS in three seconds and then returning back to 200µS in the next three seconds.
Rate starting at 100Hz, decreasing exponentially to 65 Hz and then returning to 100 Hz.
Time duration of the treatment selectable: 1 minute to 12 hours.
Low Battery Indicator: If the battery goes below 6.9 volts +/- 0.2 volts the battery symbol will flash on/off once every second.
If the battery voltage is below 6.6 (+/- 0.2) volts the unit will not turn on.
Open Electrode Detect: If an open circuit is detected at the output of channel A or B the output current will be reset at zero.
Physical dimensions: 108 x 62 x 23 mm.
Weight: 70gm without battery, 100gm with battery.
Environmental conditions for storage & transport: -10 to +50 degrees Centigrade 0-90% Humidity